Charge Time: In instrument, typically 10 hours from empty at ambient temperatures
below 30 °C. To protect the battery, charging will be terminated completely at ambient
temperatures above 40 °C. At 30 to 40 °C charging time will be prolonged. With external
charger ZG-0444 (optional accessory), typically 5 hours.
Please note: It is not recommended to charge the battery at temperatures below 0 °C
(32 °F) or over 50 °C (122 °F). Doing this will reduce battery lifetime.
4.16 Warm-up Time
Warm-up Time: < 30 s after reaching equilibrium with the ambient environment and
switching on power.
4.17 Real-time Clock
Back-up Powered Clock Drift: <0.45 seconds over a 24-hour period
4.18 CE-mark and C-Tick mark Compliance
The CE marking is the manufacturer's declaration that the product meets the
requirements of the applicable EU directives RCM mark indicates compliance with
applicable ACMA technical standards – that is, for telecommunications, radio
communications, EMC and EME
China RoHS mark indicates compliance with administrative measures on the control of
pollution caused by electronic information products according to the Ministry of
Information Industries of the People’s Republic of China
WEEE mark indicates compliance with the EU WEEE Directive
Safety
EN/IEC 61010– 1, ANSI/UL 61010–1 and CSA C22.2 No.1010.1: Safety requirements for
electrical equipment for measurement, control and laboratory use
EMC Emission
EN/IEC 61000–6–3: Generic emission standard for residential, commercial and light
industrial environments EN/IEC 61326: Electrical equipment for measurement, control
and laboratory use – EMC requirements CISPR 22: Radio disturbance characteristics of
information technology equipment. Class B Limits
IEC 61672–1, IEC 61260, IEC 60651 and IEC 60804: Instrumentation standards
Note: The above is only guaranteed using accessories listed in this document
EMC Immunity
EN/IEC 61000–6–2: Generic standard – Immunity for industrial environments
EN/IEC 61326: Electrical equipment for measurement, control and laboratory use – EMC
requirements IEC 61672–1, IEC 61260, IEC 60651 and IEC 60804: Instrumentation
standards
Note: The above is only guaranteed using accessories listed in this document
Temperature
IEC 60068–2–1 & IEC 60068–2–2: Environmental Testing. Cold and Dry Heat.
Operating Temperature: –10 to +50 °C (14 to 122 °F)
Storage Temperature: –25 to +70 °C (–13 to 158 °F)