12 Specications
CPR Targets
Category Specication
Compression Depth Target > 50 mm
( 2”)
Depth accuracy: ±10%
Compression Release Target
Force
< 2,5 kg (5,5 lbs).
Force accuracy: +1,5 kg, -2,0 kg
(+3,3 lbs, - 4,4 lbs)
Compression Rate Target 100 to 120/min ± 3/min
CPRmeter [REF 801-00140]
The CPRmeter meets the performance requirements of IEC 60601-1, 2nd
and 3rd edition.
Category Specication
Dimensions 154 mm x 64 mm x 28 mm
(6.1” x 2.5” x 1.1”)
Weight 227g (7,3 oz)
Battery 3V Lithium 123 (Li/MnO
2
)
The CPRmeter has been tested
with Energizer
®
batteries. Use
only Energizer
®
, Panasonic
®
or
Duracell
®
batteries.
microSD™ Memory Card The CPRmeter has been tested
with 2Gb microSD™ memory
cards. Always test a new card
before using it in data collection.
Temperature and Relative
Humidity
Storage temperature:
-20° to 70° C (-4° to 158° F).
Relative humidity: 0% to 75%.
Operating temperature:
0° to 40° C (32° to 104° F).
Operating atmospheric pressure :
1000 to 540 mbar (100 to 54 kPa).
Relative humidity 0% to 95%.
Sealing: Meets ISO/IEC 60529 class
IP55.
Product temperature can reach
60° C.
Electromagnetic Compatiblity Meets IEC 60601-1-2 and RTCA/
DO-160E.
Recovery time Recovery time after defibrillation: 0 s