Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030 95
Symptoms
Possible Cause(s)
Corrective Action(s)
Software
Valsalva readings
are being marked
automatically as
coughs.
When performing
Valsalva procedure,
patient is triggering a
sharp rise in pressure
resembling a cough
spike?
Activate Bristol cough marker in the
workflow configuration.
Click on Settings>Workflow
Settings>Phases. Create a copy of
the workflow. In each study phase
scroll to the Workflow Additional
Configuration Items subtitle and de-
select Auto Cough artifact
detection.
Select Auto Bristol artifact
detection and save changes to custom
Workflow. Use the new custom
workflow to complete test with the
patient.
For more information on custom
workflows refer to the Study
Management
Workflows section on page 55.
If problems continue, contact the LABORIE Service team at 1-800-333-1039 or email
service@laborie.com.
Software Error Messages
If the following error messages are displayed in Synergy software, contact the LABORIE
Service team at 1-800-333-1039 or email service@laborie.com:
Hardware Failure
Device is not ready for another process
Data could not be written to EEPROM
Data could not be read from EEPROM
SD card error
Study map full
SD card full
SD card not present
SD card faulty
Status DRAFT Effective