Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030 87
5.3.2.2 Verifying Pressure Measurement Capability
To begin verifying calibration for the PIM NXT transducers, gather one Cylinder at least 25
cm tall and one NXT Catheter. To complete calibration verification:
a) Fill a cylinder with 20cm of water.
b) Coil the NXT Catheter inside the cylinder so that all pressure ports touch the bottom
of the cylinder.
c) Connect the catheter to the relevant PIM NXT channel.
d) Click the Settings icon in the title bar, then click the System Settings > Device
Manager option. The Device Manager window will pop-up.
e) Click the Roam NXT option, then click the Run Diagnostics button. The device
diagnostics window will pop-up (Figure 69).
Figure 69: Roam NXT – Device Diagnostics
f) The Pressure should be approximately 20cm of water +/-3%. Repeat for each
channel.
NXT Go Pump Hub and Volume Transducer –
Verifying Calibration
To set up verification of calibration for the Infusion Pump NXT, located on the NXT Go Pump
Hub, and the Infusion Volume Transducer, located in the NXT Go Infusion Hook, follow the
instructions below:
1. Login to the Synergy software.
2. Ensure the Infusion Pump NXT is connected to the Synergy software.
Status DRAFT Effective