Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030 84
5.3 Verifying Calibration
LABORIE recommends users verify calibration annually, or when any module has been
shaken, dropped, or transported between sites, to confirm system accuracy.
IMPORTANT: NXT System modules do not require user calibration as the system
modules are provided calibrated. If verification procedures reveal a calibration malfunction,
contact LABORIE Service for a replacement device.
Urocap NXT – Verifying Calibration
To setup verification of calibration for the Urocap NXT, follow the steps provided below.
NOTE: To complete the calibration procedure prepare one LABORIE beaker and 500ml
of tap water.
1. Login to the Synergy software.
2. Connect the Urocap NXT to the Synergy software and set up the hardware channels.
5.3.1.1 Checking the Absolute and Relative Weight
To begin calibration verification, check the total weight measured on the Urocap NXT
without a beaker:
a) Place the Urocap NXT on a level surface and let it settle for 20-30 seconds.
b) Click the Settings icon in the title bar, then click System Settings > Device
Manager option. The Device Manager window will pop-up.
c) Click the Urocap NXT option, then click the Run Diagnostics button. The device
diagnostics window will pop-up (Figure 66).
Figure 66: Urocap NXT – Device Diagnostics
Status DRAFT Effective