Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030 81
Repeat this process, if necessary, using either a fresh wipe or the clean side of the
existing wipe until the affected area is visibly clean.
Re-insert the vent plug.
5.1.4.3 Disinfecting the Roam NXT and PIM NXT
To disinfect the Roam NXT, PIM NXT for T-DOC, and PIM NXT for Fluid, follow the
instructions provided in the Performing Manual Cleaning section on page 79 allowing the
surface to remain wet for the required contact time listed on the product label of the
cleaning solution. For any parts that may contact the patient’s skin, such as belt clips,
complete a rinse step to aid in the removal of reagent residuals. Use a clean, lint free cloth
that has been moistened with purified water to wipe the part. Repeat for a total of three
rinses.
Disinfect the area of the PIM NXT for T-DOC normally covered by the transducer doors
following the instructions below:
Remove the transducer doors and place safely aside.
Gently wipe the affected area with a cleaning wipe taking care to avoid the
transducer channels. Take care to prevent any excess cleaning reagent from
entering the transducer channels. Wipe each transducer door to remove soiling
and debris. Allow the surfaces to remain wet for the required contact time listed
on the product label of the cleaning solution.
Repeat this process, if necessary, using either a fresh wipe or the clean side of the
existing wipe until the affected area is visibly clean.
Reattach and close the transducer doors.
5.1.4.4 Disinfecting the UPP Puller NXT
Prior to beginning disinfection procedure, ensure that the UPP Puller NXT nose is separated
from the motor. Remove the Catheter Guide and the Catheter Clamp and disinfect
separately. Refer to the Assembling the UPP Puller NXT section on page 22.
To disinfect the UPP Puller NXT nose, Catheter Guide, and Catheter Clamp, follow the
instructions provided in the Performing Manual Cleaning section on page 75 allowing the
surface to remain wet for three minutes or for the required contact time listed on the
product label of the cleaning solution.
Completing Cleaning and Disinfection
Follow the instructions provided below to complete the cleaning and disinfection process.
5.1.5.1 Reconnect and Reassemble
If previously disassembled, reassemble the Roam NXT and PIM NXT by inserting
the PIM NXT into the Roam NXT collar until it is flush. Screw together re-using the
screw supplied with the device. Ensure not to overtighten the screw.
Reattach PIM NXT and Roam NXT belt clips. Refer to the Removing and Attaching
the Clip section on page 103.
If previously disassembled, reassemble the UPP Puller NXT by connecting the motor
and the nose. Reattach the Catheter Guide and the Clamp. Refer to the
Assembling the UPP Puller NXT section on page 22.
Ensure to close the pump door.