Owner's Manual
Table Of Contents
- 1 Introduction
- 2 Quick Start
- 2.1 Setting Up the Equipment
- 2.2 Gathering Supplies and Equipment
- 2.3 Preparing the Urocap NXT
- 2.4 Preparing the NXT Go Pump Hub
- 2.5 Preparing the Roam NXT
- 2.6 Preparing the UPP Puller NXT (Optional)
- 2.7 Setting Up Consumables and Preparing the Patient
- 3 How to Perform a Study
- 4 Synergy Software
- 5 Maintenance and Service
- 5.1 Cleaning and Disinfection
- 5.2 Preventive Maintenance
- 5.3 Verifying Calibration
- 5.4 Configuration
- 5.5 Computer Virus Protection
- 5.6 Disposing of Product After Use
- 5.7 Environmental Consideration of Waste Disposal
- 5.8 Backups
- 5.9 Troubleshooting Guide
- 6 Hardware
- 7 Appendix

Synergy Software
NXT Go Owner’s Manual, LBL-00145, MAN2030 57
3P Urodynamics. Alternatively create a custom workflow by adding
a UPP phase to the default workflow templates provided.
b. The Phases setting tab allows for customization of auto and manual event
markers per phase and of phase layout.
i. Refer to the Customizing Control Panel section on page 58 for
information on customizing event markers.
ii. The Phases tab provides a phase layout for the workflow selected.
Update phase layout for custom workflows by using the plus ,
up/down , and delete icons provided. Rearrange or delete
phases under the Protocol heading by clicking on the phase
heading and using the appropriate icon to complete the change.
To add a phase, Select the Protocol heading and click the plus
button. Select the desired phase from the popup window and click
OK (Figure 44).
Figure 44: Adding a Phase to a Custom Workflow
iii. Adjust Fluid Density in the Phases tab by entering a new value.
The Fluid Density setting is available in the filling, voiding, and
UPP phases. If updating Fluid Density settings, ensure to update
in all relevant phases. Fluid density can be changed during a study
using the Change Infusion Fluid button to fluid densities
previously saved to the workflow. Click the fluid density dropdown
menu to view previously saved fluid densities.
c. The Channels settings tab provides settings for hardware channels.
Refer to the Accessing Hardware Channels section on page 60.
d. The Event Summary settings tab displays events for inclusion in the
event summary. Select events for inclusion using the checkboxes
provided. View event summary in the Review and Report of the
completed study.
Select Questionnaire or Review under the Workflow Steps. Enable or disable
automatic execution under the Settings tab.
Select Report under the Workflow Steps. Customize Landscape Graph
Scaling settings by selecting scaling by Number of Pages or Time per Page.
Click the OK button to save the workflow.
4.4.2.1 Workflow Step Settings Tabs
The following settings tabs are available under the Procedure Workflow Step:
Global: Graph Settings, Channel Settings, Other Settings, EMG Settings, Profile
Settings
Status DRAFT Effective