How to Perform a Study
NXT Go Owner’s Manual, LBL-00145, MAN2030 39
Permission to Void button is selected in error, delete the event and continue
“filling phase” activities by pressing the Pump On button.
The Cystometry phase will stop 60 seconds after voiding ends. Alternatively click
the Finish Procedure button.
A Post Void Residual window will pop-up. Select the preferred method of
capturing Post Voided Residual (PVR), meaning the amount of fluid remaining in
the bladder at the completion of the voiding phase. Click OK to move to the
Review phase.
3.4.2.3 Review and Report Phases
Review study data and adjust markers in the Review phase. Click the Report button to
proceed to the Cystometry Report. Customize and save report layout.
3.5 Tests
The following tests can be completed within Urodynamic studies that provide a Filling or
Cystometry phase.
Stress Test
Commence the stress test once the patient’s bladder had been infused with 150-
200ml of fluid. Position the patient so that the meatus, the urinary opening, is
easily observed.
Select the Stress Test toggle button to begin the stress test. Clicking the Stress
Test toggle button will automatically stop the pump.
If a leak is observed select the Leak button to mark the event.
Select the Stress Test toggle button again to end the stress test. A pop-up
window will appear showing a suggested designation for the stress test.
Synergy automatically assigns each stress test based on events marked within the
range. To change the assigned type designation, select the option and press OK
(Figure 25). Changing the designation will add or remove the events marked in
the stress test.
Figure 25: Stress Test Designation Window
Repeat the test after every 150ml to 200ml infused until stress incontinence is
documented or ruled out.
Finish the selected study.