How to Perform a Study
NXT Go Owner’s Manual, LBL-00145, MAN2030 33
Click the Zero Volume button to prepare the Urocap NXT and then click the Start
Flow Phase button to start the study and move to the Flow phase.
NOTE: Pause the study by clicking the Pause button and resume the study
by clicking the resume button.
3.2.2.2 Flow Phase
Instruct the patient to void. The Synergy application will automatically start to
record data when the transducer detects the start of the flow.
NOTE: To shift the position or empty the urine beaker during the Flow Phase, click the
Change Urine Beaker button. A popup window will appear, and the study will pause.
Once the action is complete, click Continue on the popup window to resume.
Figure 20: FreeFlow Record Phase
The flow phase will stop 60 seconds after voiding ends.
A Post Voided Residual window (Figure 21) will pop-up. Select the preferred
method of capturing Post Voided Residual (PVR), meaning the amount of fluid
remaining in the bladder at the completion of the voiding phase. Use of the PVR
functions is at the discretion of the overseeing physician.
a. Calculated PVR selection automatically provides the difference between
infused volume and voided volume.
b. Drainage PVR selection allows the operator to pour drained fluid into the
Urocap NXT to measure PVR.
IMPORTANT: Do not empty the Urocap NXT beaker before adding PVR fluid
drained from patient bladder. If the beaker is filled to capacity utilize Calculated
PVR or Input PVR options.
c. Input PVR selection allows the operator to input the PVR measurement
calculated by a separate measurement device.
Status DRAFT Effective