NXT Go Owner’s Manual, LBL-00145, MAN2030 26
Figure 9: Catheters Connected to PIM NXT
Verify that the catheter appears in the consumable traceability window in
Synergy. If the catheter does not register in Synergy, disconnect the catheter
and reconnect.
Confirm positioning of the catheter using the Synergy interphase. Ask the patient
to cough to confirm the catheter’s positioning. When recording two pressures at
the same time, the cough will show equal deflection on both channels (Figure 22).
With Auto Cough detection activated, Synergy will mark these deflections as
coughs automatically; refer to Customizing Control Panel on page 58 for more
information.
Figure 10: Pressure Flow Preview Phase
IMPORTANT: T-DOC Air-Charged Catheters are intended for single-use. Synergy will
not recognize a catheter that has previously been registered in the software after the
allotted two hours to complete Urodynamic testing.
Once preferred catheter placement is confirmed, connect the Infusion Pump
Tubing to the filling lumen on the vesical catheter or filling catheter. Ensure all air
bubbles have been removed from the Infusion Pump Tubing and that the Infusion
Pump has been properly primed before connecting the line to the filling lumen on
the vesical catheter or filling catheter.
Refer to the T-DOC Air-Charged Catheters section on page 108 for information on T-DOC
catheters.
2.7.2.2 Roam NXT, PIM NXT for Fluid, and Transducer Setup
Ensure the PIM NXT for Fluid is connected to the Roam NXT. A solid green light
bar on the Roam NXT indicates that the device is ready for use. For additional