Quick Start
NXT Go Owner’s Manual, LBL-00145, MAN2030 22
2.4 Preparing the NXT Go Pump Hub
To prepare the NXT Go Pump Hub:
Place the device on a stable surface where it will not be a tripping hazard.
Connect the NXT Go Infusion Hook to the NXT Go Pump Hub.
Connect the NXT Go Pump Hub to supply mains using the NXT Go power cord and
the NXT Go power supply.
Connect the NXT Go Pump Hub to the computer using the NXT Go USB cable.
Refer to the NXT Go Pump Hub section on page 98 for connection points on the NXT GO
Pump Hub and to the Infusion Pump Tubing Setup section on page 24 to setup the pump.
2.5 Preparing the Roam NXT
Ensure the Roam NXT is charged. LABORIE recommends charging the device
overnight to ensure optimal battery power for a full day of procedures.
Confirm that the Roam NXT and PIM NXT are connected securely. The PIM NXT
and the Roam NXT must be connected to function. For information on the Roam
NXT and the PIM NXT, refer to the Roam NXT section on page 100.
2.6 Preparing the UPP Puller NXT (Optional)
Ensure the UPP Puller NXT is charged. LABORIE recommends charging the device overnight
to ensure optimal battery power for a full day of procedures. Refer to the UPP Puller NXT
section on page 107 for instructions on how to charge the device. Confirm that the UPP
Puller NXT is fully assembled and securely mounted.
Assembling the UPP Puller NXT
Identify removable parts to begin assembly of the UPP Puller NXT. The UPP Puller NXT nose
separates from the motor through a locking mechanism located at its base (Figure 3). Fit
the nose onto the motor with the locking mechanism in the unlocked position. Once the
couplers have self-aligned lock the mechanism (Figure 4). If disassembly is required,
unlock the mechanism and pull the nose away from the motor.
Figure 3: UPP Nose Locking Mechanism
Figure 4: Connect the UPP Motor and Nose
The UPP nose features the removable Catheter Guide and Catheter Clamp (Figure 5).
Attach the Catheter Guide and Catheter Clamp to the appropriate connection point. The
Catheter Guide and Catheter Clamp use magnets to connect and disconnect from the nose.
Figure 5: Left Profile – Removable Parts
Catheter Guide
Catheter Clamp
Carriage