Introduction
NXT Go Owner’s Manual, LBL-00145, MAN2030 17
To avoid risk of electric shock, do not touch the pogo pins located inside the Urocap NXT
Charging Hub or the Urocap NXT Wall Charger.
To avoid the risk of electric shock, only connect this equipment to a mains power supply
with protective earth.
This device is intended for use in a clinical environment with controlled EMC standards to
limit potential interference. The NXT System may be adversely affected by Bluetooth,
cellular, or EMC interference. Minimize interference from other Bluetooth devices by
setting up all modules of the system in proximity to each other.
Electromagnetic Interference in the form of electrostatic discharge or radiated emissions
can cause the NXT System components to temporarily disconnect. Components may
require a power cycle before operation can resume after exposure to these
electromagnetic interferences.
Only connect devices specified and approved by LABORIE to the NXT System. Connecting
any device not approved by LABORIE to the Medical Electrical (ME) system alters the
functional integrity of the product creating a safety risk.
Do not connect any multiple socket-outlets or extension cords to the NXT GO System or
any of its components
Always position the NXT System to facilitate ease of access to all electrical connections
and plugs. The device operator must be able to disconnect the device quickly and safely
in case of an emergency.
Do not alter or add to the NXT medical electrical system. Any alteration to the NXT
System by an unauthorized party transfers responsibility for meeting ME system
requirements from LABORIE to the altering party. Anyone connecting supplementary
equipment to ME equipment configures a medical system and is therefore responsible for
ensuring that the system complies with the requirements for medical electrical systems,
encompassed in the IEC/EN 60601 series.
Do not make changes or modifications to the PIM NXT for TDOC and its
radiators. Changes to intentional or unintentional radiators not expressly approved by the
party responsible for compliance could void the user's authority to operate the equipment.
Batteries are not operator replaceable (except for in the Remote Control NXT). Do not
attempt to remove the battery. All servicing of the NXT System, modules, or attachments
must be completed by LABORIE authorized personnel.
NOTE: Local laws take priority over the above-mentioned requirements and warnings;
if in doubt, consult your local LABORIE representative or the technical service department.
Status DRAFT Effective