User guide
APPENDIX
© Hermle Labortechnik GmbH Z306 Instruction manual.docx V
Table 1: Technical Data
Weight without rotor
max. speed
max. volume
max. RCF
allowable density
allowable kinetic energy
Mains power connection AC
Voltage fluctation ± 10 %
Current consumption 1,3 A
Power consumption 0,24 KW
Radio interference
Audit requirement (BGR 500) no
Environement
High
Ambient temperature
Max. relative humidity
Degree of contamination
Class of protection
EMV EN / IEC FCC Class B EN / IEC FCC Class B
61326-1
61326-1
Category B Category B
Inventory-No.:
Monitoring-Nr.:
Environement:
Maintenance contract:
30 kg
120 V / 60 Hz 1 ph
± 10 %
2,4 A
0,3 KW
13500 min
-1
4 x 100 ml
18625 x g
1,2 kg/dm
3
5595 Nm ( Rotor 221.17 V03)
230 V / 50 Hz 1 ph
for indoor use only
Use up to an altitude of 2000 m above MSL
2°C up to 35 °C
Max. relative humidity 80 % for temperatures up to 31°C,
VDE 0871, radio interference B
decreasing linearly to 50% relative humidity up to 35°C
Overvoltage category (IEC 60364-4-443) II
2
Ambient conditions (EN/IEC 61010-1)
I
Not suitable for use in hazardous environements
Interference emission, noise immunity
Noise level (depending on the rotor)
≤
63dB(A)
Write from operator
responsible service office
responsible dealer
HERMLE Labortechnik GmbH or dealer service office
Siemensstrasse 25
78564 Wehingen
Tel.: (49)7426 / 96 22-17
Fax: (49)7426 / 96 22-49