Error no.: 82Error no.: 82
Error no.: 82Error no.: 82
Error no.: 82
Cutoff power board - frequency converterCutoff power board - frequency converter
Cutoff power board - frequency converterCutoff power board - frequency converter
Cutoff power board - frequency converter
Reason: Overcurrent or undervoltage due to power supply fluctuations
Action: Restart the unit, take care that the power supply is stable
Error no.: 83Error no.: 83
Error no.: 83Error no.: 83
Error no.: 83
Preset speed cannot be reachedPreset speed cannot be reached
Preset speed cannot be reachedPreset speed cannot be reached
Preset speed cannot be reached
Reason: Preset speed cannot be reached
Action: Call service
Error no.: 90Error no.: 90
Error no.: 90Error no.: 90
Error no.: 90
Emergency lid releaseEmergency lid release
Emergency lid releaseEmergency lid release
Emergency lid release
Reason: Lid has been opened during centrifuge run.
Action: Close the lid. Danger of accident!
Only open centrifuge lid when rotor stands still
Reason: Control switch of lid lock defective
Action: Call service