User Manual
19
Error no.: 25Error no.: 25
Error no.: 25Error no.: 25
Error no.: 25
Power failurePower failure
Power failurePower failure
Power failure
Reason: Power failure while rotor is in motion
Action: Open and reclose the lid, restart the centrifuge
Error no.: 36Error no.: 36
Error no.: 36Error no.: 36
Error no.: 36
Relay for the frequency converter cannot be releasedRelay for the frequency converter cannot be released
Relay for the frequency converter cannot be releasedRelay for the frequency converter cannot be released
Relay for the frequency converter cannot be released
Reason: Defect on the power board
Action: Call service
Reason: Centrifuge lid is jamming
Action: Open the centrifuge lid manually. Grease the lid lock lightly.
If this failure occurs again, call service.
Error no.:50,51Error no.:50,51
Error no.:50,51Error no.:50,51
Error no.:50,51
Memory failureMemory failure
Memory failureMemory failure
Memory failure
Reason: Internal or external memory failure
Action: Restart the unit, if the failure occurs again, call service
Error no.: 55Error no.: 55
Error no.: 55Error no.: 55
Error no.: 55
OverspeedOverspeed
OverspeedOverspeed
Overspeed
Reason: Overspeed sensor or engine speed sensor defective
Action: Call service
Error no.: 60Error no.: 60
Error no.: 60Error no.: 60
Error no.: 60
Engine speed sensor signal is missingEngine speed sensor signal is missing
Engine speed sensor signal is missingEngine speed sensor signal is missing
Engine speed sensor signal is missing
Reason: Engine speed sensor defective or parting of a cable at the sensor
Action: Call service