Owner's manual

Page 3 of 24 INSTRUCTION MANUAL Z 206 A © HERMLE Labortechnik GmbH
© HERMLE Labortechnik GmbH 2004/01
1 PRODUCT DESCRIPTION
1.1 Usage in accordance with safety standards
1.1.1 General information
This unit is a medicine product according to the IVD guide line 98/79/EG.
1.1.1.1 Hazards and precautions
Before setting the centrifuge into operation, please read this instruction manual carefully!
For your personal safety, please review the following precautions:
• The HERMLE Z 206 A is not explosion-proof and must therefore not be operated in explosion-
endangered areas or locations. During centrifugation, it is prohibited to stay within the safety zone of
30 cm around the centrifuge or deposit hazardous substances within this area.
Centrifugation of flammable, explosive and radioactive substances, which chemically
react with high energy, is strictly prohibited!
Never spin toxic or pathogenic material without adequate safety precautions, i.e. centrifugation of
buckets / tubes without or with defective hermetic sealings is strictly prohibited.
The user is obligated to perform appropriate disinfection procedures in the case that dangerous
substances have contaminated the centrifuge and / or the accessories. When centrifuging infectious
substances, always pay attention to the General Laboratory Precautions. If necessary, contact your
safety officer.
It is prohibited to run the centrifuge with rotors that are not listed for this unit.
Under no circumstances open the lid of the centrifuge while the rotor is still running or rotating with a
speed of > 2 m/s.
Following rules must strictly be adhered to:
Do not operate the centrifuge in the case that it is not installed correctly.
Do not operate the centrifuge when dismounted (e.g. without housing).
Do not run the centrifuge when mechanical or electrical components have been tampered with
by unauthorized personnel.
Do not use accessories such as rotors and buckets, which are not exclusively approved by
HERMLE Labortechnik GmbH, except commercially available centrifuge tubes made of glass
or plastic.
Do not spin extremely corrosive substances, as they may cause material damages and impair
mechanical resistance.
Do not operate the centrifuge with rotors or buckets, which show any signs of corrosion or mechanical
damage.