Owner's manual

Page 24 of 24 INSTRUCTION MANUAL Z 206 A © HERMLE Labortechnik GmbH
4 TROUBLE SHOOTING
4.2.2 Description of the error message system
The error message is shown in the "time" display through a two-digit number (19). At the same time the
word “error” (20) is indicated in the display (see photo 20).
Photo 20
4.2.3 Error messages
Errors that may be indicated in the LCD display:
Error No.: Description
01 Rotor Imbalance
02 Imbalance sensor is defective
14 Leap of speed is too big between measurements
30 Motor Error
33 Open lid while the motor is running
34 Lid contact defective
55 Overspeed
60 Undervoltage in the intermediate circuit
70 Blackout of the relay
19