INSTRUCTION MANUAL FOR THE Z100-A CLINICAL CENTRIFUGE C0100-A January 2005
INDEX PAGE 1. General Information ........................................................................................ 1.1 Description ............................................................................................ 1.2 Safety precautions to be observed before operating the centrifuge ...... 1.3 Technical data........................................................................................ 1.4 Accessories .............................................................................
1. General Information - This manual is intended for use with the Z100-A (C0100-A) Centrifuge. 1.1 Description The Z100-A is a small bench top centrifuge designed for sedimentation of various samples. The Z100-A has a built in angle rotor capable of running up to to 4,000 rpm (1,900 x g.) This product is for research only. 1.2 Safety precautions to be observed before operating the centrifuge ! Do not operate if: - The centrifuge has not been installed correctly. - The centrifuge is partly dismantled.
1.4 1.5 Accessories Accessories supplied with each centrifuge unit: - One 6 x 15ml rotor - One Instruction manual - One Warranty card - Six Rubber Cushions Aditional accessories available: C0200-17A - Combination adapters for 5ml and 7ml tubes, (6pk.) C0230-17A - Combination adapters for 5ml and 7ml tubes, (4pk.) Warranty This centrifuge has been subject to thorough testing and quality control.
2.4 Centrifuge Installation 1. Make certain that the timer is set for the "off" position. Before operating the centrifuge, check that the power source corresponds to that on the manufacturer's rating label, then connect the power cord to a properly grounded outlet ATTENTION: Ensure that the timer is in the "OFF" position, before connecting the power cord. 3. Operation ! ATTENTION: Never operate the centrifuge if the rotor shows any signs of corrosion or mechanical damage.
4. Proper loading of the rotor 4.1 The rotor has been designed to handle samples with a maximum density of 1.2g/ml. Tubes should not be run with any substance having a density above 1.2g/ml. 4.2 When loading the rotor you should proceed according to sketch A. Fill the tubes equally by eye and insert them into the tube shields of the rotor. The difference in weight between the tubes should not exceed 2 - 3 grams. A partially loaded rotor may be centrifuged.
5. Service and Maintenance 5.1 Service of centrifuge Ther is no routine maintenance or service required on the centrifuge other than occasional cleaning. 5.2 General cleaning of the centrifuge Always keep the centrifuge housing, rotor, and rotor accessories clean by wiping down with a damp cloth. Do not use harsh cleaners or detergants on the lid or housing surfaces as it may affect the plastic. After cleaning, ensure that all parts are dried thoroughly by hand or in a warm-air-cabinet (max. temp.
6. Troubleshooting guide Centrifuge will not start Possible reason: Solution: No power supply. Check main power supply and centrifuge fuse. Possible reason: Solution: Blown fuse. Check fuse and replace if necessary (see 5.5) Lid cannot be released: Possible reason: Solution: Lid lock is jammed. Push front of lid down while pushing release button. Centrifuge cannot be started, although power is on: Possible reason: Solution: Lid not closed correctly. Close lid correctly, push down on front of lid.
7. Chart for determining g-values See section 1.3 for the correct radius of your rotor. Note the use of tube adapters will change this radius.. To determine the relative centrifugal force (RCF - value).