Instruction Manual

7
6. Troubleshooting Guide
Please refer to this guide before calling for service.
Centrifuge will not start
Possible reason: No power supply
Solution: Check that power is being supplied to the outlet
Check that the power cord is plugged into both
the wall outlet and the back of the centrifuge
Check that power cord is not damaged
Possible reason: Blown fuse
Solution: Check fuse and replace if necessary
Lid lock will not release
Possible reason: Defective lid lock
Solution: Open manually and have unit serviced
Possible reason: No power from PC board
Solution: Call for service
Possible reason: Lid lock is jammed
Solution: Call for service
Possible reason: Centrifuge is not receiving power
Solution: See “Centrifuge will not start”
Centrifuge cannot be started, although power is on
Possible reason: Lid not closed correctly
Solution Close lid correctly
Possible reason: No speed or time has been selected
Solution: Set speed and/or time
bAL: Error Message Indicates imbalance
Possible reason: Tubes not inserted symmetrically in rotor holes
Solution: Load tubes symetically
(see Section 3.4 on Loading the rotor
Possible reason: Sample liquid in tubes not balanced
Solution: Make sure the same volume of liquid is in each
tube
Possible reason: Defective or improperly adjusted balance sensor
Solution: Call service
Lid: (Error message) Lid not closed
Possible reason: Lid not closed completely
Solution: Close lid
Possible reason: Lid lock or sensor defective
Solution: Call service
Other Error Messages Er on Display
Solution: Press time or speed knob to clear error
Call service