User Manual
2.2 Required space
The centrifuge should be i
n
stalled on a rigid, even surf
ace such as a stable laboratory
bench, countertop, etc. To guarantee sufficient ventilation, ensure that the centrifuge has
at least 15cm (6 inches) of free space on all sides, including the rear.
The centrifuge should not be located in areas subject to excessive heat such as in direct
sunlight or near radiators or the exhaust of a compressor, as a buildup of heat may occur
within the chamber.
2.3 Installing the Centrifuge
Before operating the centrifuge, check that the power source (electrical outlet on the wall)
corresponds to that on the manufacturer’s rating label, then connect the power cord to the
centrifuge and the power source.
3. Installation of rotors and rotor maintenance
3.1 Rotors and accessories
The following accessories are included or available for the Prism:
Angle rotor for 24 x 1.5ml tubes (Included)
Order no. C2400-2
Tube measurement 1.5ml / 2.0ml (10 x 40mm)
Max. speed 15,000rpm
Centrifuging radius 8.4cm
Max. RCF (g-value) 21,200 x g
Rotor Wrench (Included)
Order no. C0240-RT
Rotor Lid (Included)
Order no. C0240-RC
Rotor Securing Screw (Included)
Order no. C0240-RSS
Strip Spin Adapter
Order no. C2400-SS
Tubes accepted 0.2ml tubes or 8 x 0.2ml strips
Max Speed 15,000 rpm
Range Centrifuge radius 4.32cm to 5.33cm
Range RCF (g value) 10,874 x g to 13,405 x g