User Guide

Safety Notes
WARNING! Risk of injury
CAUTION: Risk of damage to the device or errors in
pipetting of liquids.
Before starting the work with the pipet controller every
user should read these operating instructions carefully.
CAUTION:
Using the device inconsistently with the operating
instructions may result in damaging the device.
The device should be serviced only at an authorized
service center, otherwise the manufacturer will be
relieved from any liability under the warranty.
Only original spare parts and accessories, recommended
by the manufacturer, shall be used.
Only the original charger, supplied by the manufacturer,
shall be used for charging the batteries.
In case of incorrect functioning of the pipet controller,
work shall be stopped. The device shall be cleaned and
sent for repair to an authorized service center.
In the case of mechanical damage to the casing, the
device shall be immediately sent for repair to an
authorized service center.
The use of excessive force during work shall be avoided.
WARNING!
During the work with the pipet controller general safety
regulations regarding risks related with laboratory work
should be observed. Protective clothing, goggles, and
gloves should be worn.
The pipet controller shall not be used for measuring
substances with vapors of which damage the following
plastics: PP, SI, EPDM, POM.
The pipet controller shall not be used in an environment
where explosion risk is present.
Flammable liquids shall not be measured – in particular
substances with ash-point below 0°C (ether, acetone).
The pipet controller shall not be used for drawing acids
with a concentration above 1 mol/L.
The pipet controller shall not be used for drawing
solutions with a temperature above 50°C.
The pipet controller may work in temperature range
from +10°C to +35°C.
In the case of mechanical damage to the casing, the
device shall be immediately sent for repair to an
authorized service center.