Manual

15
2 Error Messages
No.
Error Message
Possible Reasons
1
Error Message when running Please contact the
distributor withAD1213 failure
The microchip can not communicate with the
AD1213. AD1213 broken or circuit board trouble.
2
Error Message when running Please contact the
distributor. with “Heatsink temperature is
too high
Heat sink temp. higher than 70℃,the
temperature sensor is broken, there is a short
circuit or the fan is not working.
3
Error Message when runningPlease contact the
distributor.” and "Heatsink sensor is
disconnected."
Heat sink temp. less than 5℃,the sensor broken,
there is a short circuit or the fan is not working.
4
Error Message when running "Please contact the
distributor."and "Temperature sensor No.1 is
disconnected."
Block sensor No.1 reports an open circuit
5
Error Message when running "Please contact the
distributor."and "Temperature sensor No.1 is
short-circuited."
Block sensor No.1 reports a short circuit
6
Error Message when running "Please contact the
distributor."and "Hotlid sensor is
disconnected"
Hot-lid sensor reports an open circuit
7
Error Message when running "Please contact the
distributor."and "Hotlid sensor is short-
circuited."
Hot-lid sensor reports a short circuit
Note:
During the warranty term, please contact Labnet or your distributor. If the instrument is opened
by an unauthorized individual the warranty is void.