FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.) 11-11-30 16:58 31 Mayfield Avenue, Edison - NJ 08837 - USA Phone: +1-732-417-0700 email: Labnetinfo@corning.com www.labnetinternational.
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.) 11-11-30 16:58 TABLE OF CONTENTS 1 - INTRODUCTION .............................................................................................................. 2 2 - WORK SAFETY INSTRUCTIONS ..................................................................................... 2 3 - LIMITATIONS OF USE .....................................................................................................
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.) 11-11-30 16:58 1 - INTRODUCTION The FASTPETTE PRO pipette controller is a device intended for pipetting liquids with the use of measuring pipettes. It can work with all types of glass or plastic pipettes in the volume range from 0.5 ml to 100 ml. The FASTPETTE PRO pipette controller is a medical device intended for professional use, in vitro diagnostics, and testing samples collected from the human body.
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.) 11-11-30 16:58 3 - LIMITATIONS OF USE • The FASTPETTE PRO may not be used for measuring substances, the vapours of which damage the following plastics: PP, SI, EPDM, POM. • The pipette controller may not be used in an environment where explosion risk is present. • Flammable liquids shall not be measured - in particular substances with flash-point below 0 °C (ether, acetone).
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.) 11-11-30 16:58 Filling the pipette Before aspirating is started, set the speed by successive pressing the SPEED switch (fig. 1C) until the display shows the right speed (fig. 1E). • HIGH speed - fast aspirating (fig. 2.1) • LOW speed - slow aspirating (fig. 2.2) It is recommended to choose the LOW position for pipettes with a volume up to 5 ml, and the HIGH position for pipettes with a volume greater than 5 ml.
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.) 11-11-30 16:58 6 - TROUBLESHOOTING If during your work the functioning of the pipette controller is incorrect, check the cause and remove the fault. Problem Possible cause The pipette falls out (the The pipette holder is dirty holding force of the pipette or wet (fig. 1G). is to small), or tilts to the side too much. Action Take out the pipette holder, clean, wash and dry.
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.) 11-11-30 16:58 7 - REPLACING THE FILTER CAUTION: The work safety instructions given in section 2 shall be observed when disassembling the pipette controller. The filter replacement is necessary, if drawing efficiency deterioration is observed. The direct reason may be dirty filter after a long period of use. The procedure is shown in (fig. 4). • Remove the pipette. • Unscrew the nosepiece (fig. 4A).
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.) 11-11-30 16:58 WARNING! If disposable batteries are used, it is unacceptable to connect the device to the charger. In order to prolong the life span of the rechargeable batteries, the following rules should be followed: 1. Before the pipette controller is activated for the first time, the batteries should be charged. 2. FASTPETTE PRO should not be left uncharged for a longer period of time. 3.
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.) 11-11-30 16:58 11 - ORDERING INFORMATION The FASTPETTE PRO comes with a universal charger and a set of adapters in different versions: EU, US, UK and Australia. Choose your country used adapter and connect to the housing. To mount the adapter, it should be inserted into the slots of the housing in the direction of the arrow (1), until you hear a click engagement.
FASTPETTE PRO INSTRUKCJA (474P.) Nr rys. 015.9001.00.07_FASTPETTE PRO INSTRUKCJA (474P.