Labnet Incubator Shaker I5311-DS and I-5311-DS-230V Instruction Manual PO Box 841 Woodbridge, NJ 07095 USA Phone: 732-417-0700 Fax: 732-417-1750 www.labnetlink.
CONTENTS OF THIS USERS MANUAL Graphic Symbols Safety Precautions Introduction Specifications Unpacking Installation page page page page page page 1 1 2 2 2 3 Controls & Calibration Operating the I-5311-DS Care and Maintenance Troubleshooting Guide Technical Support / Service Accessories page page page page page page 3 5 5 6 6 7 GRAPHIC SYMBOLS Your Labnet I-5311-DS incubator shaker uses internationally accepted graphic symbols to help convey information to the user and to call the users attention to i
INTRODUCTION This manual covers the specifications, operation and use of the Labnet I-5311-DS shaking incubator and its accessories. Please pay special attention to the Safety Precautions section in this manual. The I-5311-DS shaking incubator from Labnet International provides an extremely stable temperature environment, and an aggressive shaking motion for the mixing of materials for cell culture growth at ambient and above ambient temperatures.
The package should include: I-5311-DS Shaking Incubator 1 full shelf 2 shelf brackets 4 adjustable feet Instruction manual Power cord (both EU and UK cords in 230V models) Warranty Card Complete and return the warranty card to register your new I-5311-DS Shaking Incubator INSTALLATION Install the four adjustable feet and locate Shaking Incubator on a stable, flat, very solid surface near a grounded electrical outlet.
-Heating Indicator-This indicator will illuminate GREEN when the controller is calling for heat from the heater. This indicator will be on continuously while the oven heats up to the set temperature and will then cycle on and off at the set temperature. -Safety Thermostat-The safety thermostat located on the back of the unit is manually set and completely independant of the Main Termperature Controller.
OPERATING THE LABNET I-5311-DS INCUBATOR SHAKER As advised in the installation section, it is very important that the unit be level and placed on a very solid surface. When loading the platforms, it is also important to load these in a balanced manner or the unit can vibrate and possibly walk. Shaking unbalanced loads for extensive periods may also damage the shaker.
TROUBLESHOOTING GUIDE PROBLEM Oven will not power up or not heat POSSIBLE SOLUTION Check power cord, outlet and unit circuit breaker Shaking platform will not operate Check timer set to Hld or a time setting Check that nothing is blocking shaker platform Press shaker STOP/START button Temperature too high Check set point and readjust if necessary Check calibration Chamber temperature goes above set point and settles back to set point Normal operation in initial heat up or if door opened for a long pe
ACCESSORIES I-5322 I-5323 I-5330 I-5331 I-5330-50 I-5330-125 I-5330-250 I-5330-500 I-5330-1000 I-5330-2000 Full shelf w / brackets Half shelf w / brackets Universal flask clamp platform Flat mat platform 30 x 50ml flasks platform 20 x 125ml flasks platform 12 x 250ml flasks platform 8 x 500ml flasks platform 6 x 1000ml flasks platform 4 x 2000ml flasks platform Flask clamps are available in 50ml, 125ml, 150ml, 500ml, 1000ml, and 2000ml. Test tube rack clamps are also available.