5.2 DESCRIPTION OF POSSIBLE ERRORS
E 11: This error message appears, when something is wrong with motor regulation (PWM
regulator, pulse generator, motor). Shaking incubator is automatically stopped. Call
nearest service.
E 12: This error appears, when the motor does not reach set speed in 30 seconds.
Shaking incubator is automatically stopped. Call nearest service.
E 13: This error appears, when set speed oscillates for more than 100 RPM in 2 seconds.
Shaking incubator is automatically stopped. Call nearest service.
E 21: Temperature sensor does not work properly regarding to heater. Shaking incubator
is automatically stopped. Call nearest service.
E 22: Temperature sensor notices 5°C higher temperature than set temperature. Shaking
incubator is automatically stopped. Call nearest service.
E 23: Set temperature is not reached in 2 hours. Shaking incubator is automatically
stopped. Call nearest service.
6.0 TECHNICAL SPECIFICATIONS
6.1 TECHNICAL DATA
Heater power 700 W
Fan power 17 W
Shaking motor power 35 W
Fuse 2 x 3.15A 250V
2 x 6.3A 115V
Environment temperature 4°C to 65°C
Relative humidity Up to 85%, non condensing
Speed RPM regulation Digital, load independent,
from 20 to 300 RPM in 1 RPM steps
Shaking orbit 19 mm
Temperature operating range 5°C above room temperature to 70.0°C
Temperature sensor PT100
Temperature accuracy ±0.5°C
Temperature stability ±0.5°C
Timer 1 min - 99 hours, timer HOLD function
Capacity 16 x 125 ml, 9 x 250 ml, 5 x 500 ml, 4 x 1000 ml
Dimensions W x D x H 370 mm x 530 mm x 400 mm
Weight 20.2 kg
Overvoltage category Installation category I