222DS Benchtop Shaking Incubator User Manual I-5222-DS I-5222-D-230V
BENCHTOP SHAKING INCUBATOR 222DS USER'S MANUAL TABLE OF CONTENTS Page 1.0. GENERAL INFORMATION 1.1 DEFINITION 1.2 KEY FEATURES 3 3 3 2.0. TECHNICAL DESCRIPTION 2.1 CONSTRUCTION 3 3 3.0 INSTALLATION 3.1 UNPACKING 3.2 SELECTING THE RIGHT PLACE 3.3 ATTACHING POWER CORD 3.4 ATTACHING SHAKER PLATFORMS 3.5 STARTING UP 3.5.1 STARTING UP CHECK LIST 3.6 GENERAL SAFETY RECOMMENDATIONS 4 4 4 5 5 5 5 5 4.0 INSTRUCTION FOR USE 4.1 INTRODUCTION 4.2 BASIC OPERATIONS 4.
BENCHTOP SHAKING INCUBATOR 222DS 1.0 GENERAL INFORMATION 1.1 DEFINITION USER'S MANUAL 222DS BENCHTOP SHAKING INCUBATOR is a combination product of a shaker and a temperature chamber, designed for shaking and heat treatment of sensitive samples. It replaces two devices, reduces time and space needed. It is very suitable for biochemistry, microbiology and medicine laboratories in which thermal and shaking treatment of samples is a daily used routine. 1.
BENCHTOP SHAKING INCUBATOR 222DS 3.0 INSTALLATION 3.1 UNPACKING USER'S MANUAL Before starting the installation, carefully examine the delivery for possible damage or missing parts: - - - Open the box and lift the device, together with the shock absorbers, out of the box. Remove the shock absorbers and check that the machine has not been visibly damaged during the transportation. Keep the packing material until you are sure that the machine functions properly.
BENCHTOP SHAKING INCUBATOR 222DS 3.3 USER'S MANUAL ATTACHING POWER CORD Fit the mains cord, which is included in the delivery, into the mains receptacle. Connect the other end of the cord to a grounded wall socket. 3.4 ATTACHING SHAKING PLATFORMS The SHAKING INCUBATOR has a variety of platforms available, to meet most of the shaking needs. These platforms mount to the shakers via four mounting platforms which easily plug into four rubber mounting points on the top of the shaking incubator. 3.
BENCHTOP SHAKING INCUBATOR 222DS 4.0 INSTRUCTIONS FOR USE 4.1 INTRODUCTION USER'S MANUAL POWER 1. POWER switch - switch ON (power switch shines) or OFF the shaking incubator. 2. RUN green signal light - shines when the shaking incubator is running. 3. TEMPERATURE yellow signal light - shines when the shaking incubator is set on temperature. 4. SPEED yellow signal light - shines when the shaking incubator is set on speed. 5. TIME yellow signal light - shines when the shaking incubator is set on time.
BENCHTOP SHAKING INCUBATOR 222DS USER'S MANUAL Press and hold encoder knob for more than 2 seconds. Temperature signal light blinks. With rotating encoder clockwise (+) or counter-clockwise (-), set temperature to desired value: 37.5 37.5°C POWER If you want to use the shaking incubator without temperature regulation - set OFF on display. You get OFF function on display, when you rotate encoder under 0.0°C or above 70.0°C. POWER Press encoder knob. Speed signal light blinks.
BENCHTOP SHAKING INCUBATOR 222DS USER'S MANUAL If you want to use the shaking incubator without shaking - set OFF on display. You get OFF function on display, when you rotate encoder under 20 RPM or above 300 RPM. Press encoder knob. Time signal light blinks. With rotating encoder clockwise (+) or counter-clockwise (-), set time to desired value from 1 min to 99 h: 70.3 70h 30 min 1.55 1h 55 min 0.43 43 min 0.
BENCHTOP SHAKING INCUBATOR 222DS 4.3 USER'S MANUAL CHANGING PARAMETERS DURING OPERATION Note: You can not modify time during operation. If you want to change Temperature during operation, press and hold encoder knob for more than 2 sec. Temperature signal light blinks. Rotate encoder clockwise (+) or counter-clockwise (-), to set the desired value. In the meantime, temperature signal light pulses. When you set the new value, press START/STOP key only ONCE.
BENCHTOP SHAKING INCUBATOR 222DS 5.2 USER'S MANUAL DESCRIPTION OF POSSIBLE ERRORS E 11: This error message appears, when something is wrong with motor regulation (PWM regulator, pulse generator, motor). Shaking incubator is automatically stopped. Call nearest service. E 12: This error appears, when the motor does not reach set speed in 30 seconds. Shaking incubator is automatically stopped. Call nearest service. E 13: This error appears, when set speed oscillates for more than 100 RPM in 2 seconds.
BENCHTOP SHAKING INCUBATOR 222DS 7.0 USER'S MANUAL TEMPERATURE RE-ADJUSTMENT Shaking incubator has a software for temperature re-adjustment. First you have to measure the temperature in the middle of the chamber, or even better in the sample, for minimum 2 hours. You have to measure the temperature with a calibrated digital temperature thermometer, with precision 0.1°C or more.
BENCHTOP SHAKING INCUBATOR 222DS USER'S MANUAL Cleaning: • For cleaning the lid, do not use the cleaners, which could damage the surface! Do not use: acetone, benzene, paint thinner, carbon tetrachloride or other aggressive liquids, because they can damage the surface of acrylic lid. • Do not use detergents containing ketones, esters and aromatics. • Do not use cleaners with solid and abrasive particles. • Do not use polishes.
BENCHTOP SHAKING INCUBATOR 222DS 9.0 USER'S MANUAL POSITIONING OF ATTACHMENTS ON SHAKING PLATFORM You can mount different attachments on shaking platform. Follow below diagram as a guideline for positioning different flasks and accessories. EQUIPMENT DISPOSAL This equipment is marked with the crossed out wheeled bin symbol to indicate that this equipment must not be disposed of with unsorted waste.
Declaration of Conformity Number: CE 00701 Labnet International, Inc,., 31 Mayfield Ave., Edison, NJ 08837 USA Labnet International declares that the devices described below are in conformity with the EC directives listed. In the event of unauthorized modification of any of the devices listed below, this declaration becomes invalid.
LIMITED WARRANTY Labnet International, Inc. warrants that this product will be free from defects in material and workmanship for a period of two (2) yeasr from date of purchase. This warranty is valid only if the product is used for its intended purpose and within the guidelines specified in the supplied instruction manual. Should this product require service, contact Labnet International, Inc.’s Service department at 732-417-0700 to receive a return authorization number and shipping instructions.