Operation
1. Connect the water bath to an alternating current supply of voltage specified on the rating
plate.
2. Ensure the water level is above the elements. On boiling baths connect the constant level
devicetoawatersupplyanddrain.
DANGER:THISWATERBATHMUSTNOTBECONNECTEDTOADIRECTPOWERSUPPLY
3. Switch the MAINS on then Turn the simmerstat/thermostat or temperature controller up as
requiredandchecktheindicatinglampisilluminated.
4. Anindicatinglampnexttothesimmerstatwillilluminatewhentheheatersareoperating.
5. Ensure the over heat cut off is set by pressing it in or the overtemperature alarm AL4 is set
abovethedesiredoperatingtemperature(dependingonwhichispresent)
SafetyCut/Overtemperature
If the water bath is fitted with a manual reset button, be sure this is depressed before use. If the
water level is too low or runs out the switch will pop out and turn the heaters off until the reset is
depressed.
Ifthewaterbathisfittedwithaninbuiltoverheatsafetythermostat.Itmustbesettoslightlyabove
thedesiredsetpointtemperatureandwillpreventoverheating.ItwillmaintaintheAlarmSetValue.
ChangetheALM4valuetoslightlyabovethesetpointtemperatureassetoutintheinstruction
page.TheAlarmwillbeactivatedifthebathexceedstheALM4valueorifthesensorisbrokenor
damaged.Torestartthebathtoheat,acknowledgethealarmbypressingthepageandscrollkeys
simultaneously(lefttwokeysonkeypad).
Maintenance
The casing is finished with stainless steel and to maintain appearance should be wiped over with a
nonabrasivecleanser.
Theinteriorisalsostainlesssteelandmayalsobecleanedwithasolvent
LaboratoryEquipmentPtyLtdInstructionManual–CalibrationBottleTestingBathPage3