Operation
1. Connect the heating block to an alternating current supply of voltage specified on the rating plate
mountedonthesideorrearofthechamber.
DANGER:THISUNITMUSTNOTBECONNECTEDTODIRECTCURRENTSUPPLY
1. Turnonthemainsandchecktoseethattheilluminatinglampisilluminated.
2. Set the target temperature or set point temperature using the up and down arrows on the front of
thecontroller.
3. Final adjustment of the temperature controller may be required after the heating block has
reached operating temperature and this should be checked with a suitable thermometer located
intheheatingblock(Athermometerisnotsuppliedwiththeunit).
TroubleShooting
SYMPTOM
REMEDY
NoPower
(IndicatorLightisoff)
1. Check heating block is plugged in and
powerswitchedon.
2. Ensure mains power supply point is
functioning by using a test appliance on
powersocket.
3. ChecktheinternalRCDhasnottripped.
Failuretoheatormaintaintemperature
(Indicatorlightison)
1. Checkthetimerissetcorrectly.
2. Checkthetimerisreset(turnoff/on)
3. Ensure the temperature controller set
pointisaboveambient.
4. Check the safety controller fitted is above
themaincontrollersetting.
If the fault cannot be found call your distributor or the manufacturer quoting the serial number of the
unitfromthemanufacturerslabel.
3
LaboratoryEquipmentPtyLtdInstructionManual–HeatingBlock