IN STRUC TION MANUA L Laboratory Equipment Pty Ltd email: sales@labec.com.au Ph: 02 9560 2811 • Fax: 02 9560 6131 www.labec.com.
Manual No.: J-NBT LABEC Tissue float bath Operation Manual Model : J-NBT Please read this Operation Manual carefully before use.
Operation Manual Table of Contents Use of Operation Manual !!!!!!!!!!!!!!!!!!!!!!!!! 3 T o S U P E R V I S O R ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! 4 SAFETY SYMBOLS !!!!!!!!!!!!!!!!!!!!!!!!!!!! 5 Safety Precautions !!!!!!!!!!!!!!!!!!!!!!!!!!!! 6 1. Intended use of the Apparatus !!!!!!!!!!!!!!!!!! 8 2. Feature of product !!!!!!!!!!!!!!!!!!!!!!!! 8 3. Appearance !!!!!!!!!!!!!!!!!!!!!!!!!!!!! 8 4. Installation !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! 9 5. Name & functions of the controller !!!!!!!!!!!!!!!!! 10 6.
Operation Manual Use of Operation Manual ! Please read through and understand this Operation Manual before operating the product. After reading, always keep the manual nearby so that you may refer to it as needed. When moving the product to another location, be sure to bring the manual as well. ! If you find any incorrectly arranged or missing pages in this manual, they will be replaced. If the manual it gets lost or soiled, a new Operation Manual can be purchased.
Operation Manual To USER in charge of this equipment. ! Help ! Keep the operator in understanding this manual before use in operation. this manual near the machine for easy access by the operator. J-NBT-M-E-09-001(REV.
Operation Manual SAFETY SYMBOLS This operation manual and this product use the following safety symbols. Please keep in mind the meaning of each of the symbols to ensure safe use of the product and to prevent any damages, accidents, or problems resulting from negligence of these safety symbols. (As using symbols may depend on the product, not all of the symbols may be used.) SAFETY SYMBOLS ! Indicates the possibility of personal injury or death. Never fail to follow the operating procedure.
Operation Manual Safety Precautions The following safety precautions must be observed and followed to avoid fire hazard, electrical shock, accidents, and other failures. Keep them in mind and make sure that all the precautions are observed and followed properly. Laboratory Equipment Pty Ltd assumes no liability or any responsibility against any damages, accidents, or problems resulting from negligence by the users.
Operation Manual Installation ! When installing products be sure to observe Installation Precaution described in this manual. ! To avoid electrical shock, connect the protective ground terminal to electrical ground (safety ground). ! When applying power to the products from a switchboard, be sure work is performed by a qualified and licensed electrician or is conducted under the direction of such a person. ! Be sure to use the AC power cable provided.
Operation Manual 1. Intended use of the apparatus. The tissue float bath is used for removing paraffin wrinkles and foam(bubble) in pathology room & hospital laboratory, is made of aluminum casting. 2. Feature of product. ! Digital controller regulates and ensures precision temperature control. ! Digital setting /readout display. 3. Appearance. " Bath # Increase value key : Used to change setting value. $ Decrease value key : Used to change setting value. % Setting key : Used for setting up temp.
Operation Manual 4. Installation. 4.1 Precautions for Installation ! This equipment is designed and manufactured to operate properly only in use of rated voltage. Refer to specification section of this manual for rated voltage before installation. ! Use AC power cable that comes with (or installed to) the product. ! Do not touch the Power Cord wet handed. ! Avoid locations where the equipment is exposed to direct sunshine. ! Do not place the equipment in a corrosive atmosphere.
Operation Manual 5.functions of controller. 5.1 Name of the controller " : Display window # : Increase Value Key $ : Decrease Value Key % : Setting key & : Power ON/OFF switch 5.2 How to set temperature. 1) Turn on power switch 2) Press key one time for 1 second. Press key and key to set required temperature. 3) When setting is finished, pressing key at full length for 2 seconds, and all setup values will be reset as set values and return to normal mode. 5.
Operation Manual 5.4 Controller Parameter Configuration Mode * Temp. Deviation Value Correction If the real temperature is 50 degree and displayed temperature is 52 degree, you have to put in the value 2 and press set Key. If the real temperature is 50 degree and displayed temperature is 48 degree, you have to put in the value -2 and press set Key. J-NBT-M-E-09-001(REV.
Operation Manual 5.5 The error message of controller and cause Parameter Configuration Mode Display Er1 Er2 Cause Temperature sensor short circuit Breaking of temperature sensor wire Display Er3 Er4 Cause The error for low temperature The error for high temperature 6. Maintenance and service. 6.1 In case of cleaning inside of the bath ! Ensure that the apparatus has cooled down to a safe temperature to touch. - Disconnect the power plug from power outlet.
Operation Manual highly flammable material or explosive material inside the bath which may cause an explosion or a fire. 6.3 When the equipment is not used for long time. - Remove power plug from power outlet. - Clean with soft clothes. - Pack the equipment in an appropriate way and store in a safe place for storage. 7.Trouble shooting. " In case apparatus fails to work at all :- ! Control unit switched off :=> Ensure unit is on.
Operation Manual 9. Specification. 9.1 Technical Specification Model Bath inner size (W x Ø) J-NBT mm Ambient+5' ~ 60' Range Temperature 220 x 60 Controller Digital Setting & Display Regulator SSR type Thermo-couple diode sensor Heater (sheath) W 200 Interior Aluminum plate with epoxy coated. Exterior Steel plate with powder coated finish Material Power source Net Weight AC 240V , 50/60Hz , 1 Phase Kg 2.75 9.2 Circuit Diagram J-NBT-M-E-09-001(REV.
Operation Manual 9.3 Assembly Drawing J-NBT-M-E-09-001(REV.