User guide
OPERATINGINSTRUCTIONS:ICB-12
Connecttheinstrumenttoagroundedmainsoutlet,afterensuringthattheVoltageisthesameasstatedonthenameplate.
Theinstrumentmustbeconnectedtoanearthed(grounded)supply.
SecuretheInstrumenttotheBathusingthefixingbow.
Useonlydistilledorsoftwater,toavoidbuild-upofscaleontheheatingelement.Donotuseinflammableliquids.
Theminimalwaterlevelmustbeabout3cmabovetheholesnearthepumphousing.
Thetwospringclampsareintendedforfixingofareferencethermometer.
Switchtheinstrumenton,andadjusttherequiredtemperaturebyturningtheknob,ifthereisadeviation,betweenthe
temperaturesofthebathandtheknob(afterthetemperatureinthebathhassettled),adjusttheknobaccordingly.
Whentheheatingelementisinaction,theredpilotlamplightsinaccordancewiththeswitchingofthethermostat..
Theflowcanbereducedbyaclamppressingaplasticorrubberhosefastenedonthejetnozzle.
Thecirculationpumpallowsalsothecirculationoftemperaturecontrolledwatertoinstrumentsoutsidethebaths.Inthese
instancesitisimperativethatthewaterisreturnedtothebath,preferablyinthevicinityoftheheatingelement.
SAFETYModel,theICB-12isequippedwithanadditionalsafetyCut-Off,preventingelementburn-out,ifthewaterlevelfalls.
Toresetpushthebrownbuttonontherearoftheinstrument.
Theinstrumentsdonotrequireanymaintenance,excepttheperiodicremovalofscaleonthepumpandheatingelement.The
pumphousingcaneasilybecleanedbyremovingthethreescrewsatthebottomofthepumphousing.Afterremovingthe
instrumentfromwater,thepumpshouldbeplaceddownwards(toavoidwaterpenetration).
SPECIFICATION
Temperaturerange:Slightlyaboveambientupto+1000CAwithexternalcooling,thiscanbeextendedtoupto1000G.
Accuracy:approx.10.500at37°C(dependingonoperatingconditions).
Voltage:220V,50Hz(or115V,50/60Hz),1000W
LaboratoryEquipmentPtyLtdInstructionManual-ImmersionThermostatPage4