GW2045 – GW1060 Manual Pag. 29 di 43
6.13. COMMUNICATION WITH PRINTER
A serial port (RS232) is available as a standard feature on the appliance, to which the SMEG printer, named
“WD-PRINTE” with code 901783, may be connected.
Once of the accessories supplied is a flat extension to connect RS232 in a more accessible position.
The printer is capable of printing out the following data:
Serial Number of the WD1050 "Serial Number"
Number of cycles carried out "Cycle N."
Date and Time
Program Number indicating Thermal Disinfection Temperature and holding time (for example 93°C
10’)
Detergent control
o "Inflow Detergents P1 Deter" - for the detergent;
o "Inflow Detergents P2 Neutr" - for the neutralizing agent;
Temperature control
oTL1’: indicates the temperature measured inside the washing chamber. It is the main
temperature of the cycle;
oTLC’: Indicates the control temperature (on hospital models, not on laboratory models).
oTarget T’: stands for the target temperature value of the phase
Heating element
oRes ON’: indicates that the heating element is on;
oRes OFF’: indicates that the heating element is off;
Starting Time of the Cycle
Process of each Phase
End of Cycle time
Eventual errors if Cycle terminates incorrectly.
A
0
’: stands for the thermal disinfection effectiveness; It measures the effectiveness of the thermal
disinfection process carried out..
GW2045 – POSITION OF PORT RS232
fig. 8 - Rear of GW2045, detail of RS232 connection, where it is possible to connect the external printer. Position Rear cross section, left side.