DO’S AND DON’T’S FOR ULTRASONIC CLEANERS SYSTEMS
DO
Use the proper cleaning solution for the jobs. Take time in selecting the right
medium for all tasks.
Use beakers and positioning covers or auxiliary pans for additional solutions.
Rinse thoroughly after each cleaning application.
Keep tank bottom free of silt and dirt accumulation.
Rack or suspend articles to be cleaned away from tank bottom and sides.
DON’T
Run machine without solution in the tank.
Use plain water in the tank, unless absolutely imperative. Always add some
wetting agent.
Use acidic or strong alkaline solutions directly in the stainless steel tank.
(Beakers).
Allow skin contact with doubtful chemicals.
Use wax and compound removers on plastic or acrylics.
Allow surgical and dental instruments to remain in Tartar, Light Stain and
Permanent Cement Remover for more than two minutes to avoid discoloration.
Place heavy items directly on the tank bottom.
SYMPTOM PROBABLE CAUSE
No activity in tank, indicating light off. Defective power cord.
No power at mains supply outlet.
No activity in the tank, indicating light on. Blown fuse.
Component failure.
Faulty time switch.
Weak cavitation. Line voltage too slow.
Fluid is dirty.
Transducer problem.
Fluid not degassed.
Fluid too hot.