User Manual
POTASSIUM
HYPOCHLORITE
SULPHUROUS
SODIUMCHLORIDE
TARTARIC
SODIUM
HYPOCHLORITE
Operation
1. Connectthebathtoanalternatingcurrentsupplyofvoltagespecifiedontheratingplate.
2. Ensuretheoillevelisabovetheelements.
DANGER:THISBATHMUSTNOTBECONNECTEDTOADIRECTPOWERSUPPLY
3. Switch the MAINS on then Turn the simmerstat/thermostat or temperature controller up as
requiredandchecktheindicatinglampisilluminated.
4. Anindicatinglamponthecontrollerwillilluminatewhentheheatersareoperating.
5. Ensure the over heat cut off is set by pressing it in or the overtemperature alarm AL4 is set
abovethedesiredoperatingtemperature(dependingonwhichispresent).
SafetyCut/Overtemperature
If the bath is fitted with a manual reset button, be sure this is depressed before use. If the level is
tooloworrunsouttheswitchwillpopoutandturntheheatersoffuntiltheresetisdepressed.
Ifthebathisfittedwithaninbuiltoverheatsafetyprotection.Itmustbesettoslightlyabovethe
desiredsetpointtemperatureandwillpreventoverheating.ItwillmaintaintheAlarmSetValue.
ChangetheAL4valuetoslightlyabovethesetpointtemperatureassetoutintheinstructionpage.
TheAlarmwillbeactivatediftheincubatorexceedstheAL4valueorifthesensorisbrokenor
damaged.AdjustingtheAlarmHysterisisvaluewillcausetheincubatortocooldownbythatvalue
beforeitbeginsheatingagain.ForexampleiftheAlarmHysterisisissetto100theincubatormust
cooldownby100°cbelowAL4beforeitwillbegintoreheat.Whereyourequiretheincubatorto
switchoffcompletelyshouldanalarmconditionresultthenpleasecontactthemanufacturerto
purchasetheOST100manualovertemperatureresetcontrolsystem.
Maintenance
The casing is finished with stainless steel and to maintain appearance should be wiped over with a
nonabrasivecleanser.
Theinteriorisalsostainlesssteelandmayalsobecleanedwithasolvent
LaboratoryEquipmentPtyLtdInstructionManual–OilBathPage