Owner's manual

Operation
1. Connectthewaterbathtoanalternatingcurrentsupplyofvoltagespecifiedontheratingplate.
2. Ensure the water level is above the elements. On boiling baths connect the constant level device to a  
watersupplyanddrain.
DANGER:THISWATERBATHMUSTNOTBECONNECTEDTOADIRECTPOWERSUPPLY
3. Switch the MAINS on then Turn the simmerstat/thermostat or temperature controller up using the up or
downkeysonlyasrequired.
4. Anindicatinglamp‘OP1’willflashwhentheheatersareoperating.
5. Ensure the manual reset over heat cut off (brown button below the controller) is set by pressing it in or that
the over temperature alarm AL4 is set above the desired operating temperature (depending on which is
present).
SafetyOvertemperaturecutoffmanualresetbutton
If the water bath is fitted with a manual reset button, be sure this is depressed before use. If the water level is too
loworrunsouttheswitchwillpopoutandturntheheatersoffuntiltheresetisdepressed.
Ifthewaterbathisfittedwithaninbuiltoverheatsafetyalarmitmustbesettoslightlyabovethedesiredsetpoint
temperatureandwillpreventoverheating.ChangetheALM4valuetoslightlyabovethesetpointtemperatureasset
outintheinstructionpage.TheAlarmwillbeactivatedifthebathreachestheALM4valueorifthesensoris
brokenordamaged.Torestartthebathtoheat,acknowledgethealarmbypressingthepageandscrollkeys
simultaneously(lefttwokeysonkeypadtogetherfor1secondonly).
Maintenance
The casing is finished with stainless steel and to maintain appearance should be wiped over with a non abrasive
cleanser.
Theinteriorisalsostainlesssteelandmayalsobecleanedwithasolvent.
TroubleShooting
SYMPTOM
REMEDY
NoPower
(IndicatorLightisoff)
1. Check the water bath is plugged in and
powerswitchedon.
2. Ensure mains power supply point is
functioning by using a test appliance on
powersocket.
3. CheckinternalRCDhasnottripped.
Failuretoheatormaintaintemperature
(Indicatorlightison)
1. Ensure the temperature controller set
pointisaboveambient.
2. Check the safety controller fitted is above
themaincontrollersetting.
3. Cutoutbuttonhaspoppedout(iffitted).
If the fault cannot be found call your distributor or the manufacturer quoting the serial number of the unit from the
manufacturerslabel.
DeclarationofConformity
Each product is thoroughly inspected and tested to not only ensure that it meets the specifications provided, but to
also meet Australian Electrical Standard AS3820 and EMC Standard AS/NZ1044:1995,and therefore being
accreditedwithaCTicklabel.
3LaboratoryEquipmentPtyLtdInstructionManualBaths2013Ver1.01