User Manual
ExternalSteamFittings
1. Springloadedsafetyvalvepresettooperateat215K.P.A.(Safetyvalve4ondiagram).
2. Automaticairvent(airvent2).
3. Manuallyoperatedpressurereleasevalveconnectedwithopening/closingmechanism(exhaustvalve).
4. Manuallyoperateddrainvalve(drain1).
5. Adjustable steam release valve. If autoclaving sealed items, the adjustable steam release valve may be
closed to prevent the items exploding or blowing off lids or seals when the autoclave has completed its
cycle. In this case it is desirable to allow the autoclave to depressurise gradually before attempting to
openthelid.Thevalvemaybepartiallyopenedtopermitagradualreleaseofsteam(dumpvalve5).
6. Pressuregauge.
Openingthelid
Step1.
Pushes handle ‘A’ towards the rear of the autoclave and hold firmly in open position to release steam
through the steam release valve. Note: As this handle is operated it contacts an electric microswitch
and turns the heaters in the autoclave off, therefore to begin a new cycle remember to repress the
startbutton.
Step2.
Turnthehandles‘BB’clockwiseapproximately60degreestoreleasethelidfromthegasket.
Step3.
Turnthehandles‘CC’anticlockwiseapproximately20degreestoreleasethelidfromthelocatinglugs.
Step4.
Holdthelidandslightlyliftupwardsasthespringsbringopenthelidusinghandles‘CC’.
Closingthelid
Step1.
Closethelidusinghandles‘CC’andturnclockwise.
Step2.
Turn handles ‘BB’ anticlockwise compressing the gasket. It may be necessary to hold ‘CC’ when
commencingtoturn‘BB’
Step3.
Pull the handle ‘A’ until the locking pin is in position and the microswitch is activated. This action will also
closethepressurereleasevalve.
InternalTemperatureProbes(Optionalnotonallmodels)
If the autoclave is fitted with two internal temperature probes. To change these probes you must remove the
side panel and unscrew the glands in which they enter the chamber. The probes are ‘K’ type thermocouple
autoclave type and are connected to an external plug for monitoring by chart recorder or computer or P.L.C.
Included is two blank plugs which you can screw into the vessel if you do not wish to use the internal
probes be sure you use plumbers tape before screwing theses into the vessel or replacing them in the
vessel.
4LaboratoryEquipmentPtyLtdInstructionManualAutoclaveAA13,16,18,20Ver2.512013