User Manual

thelocalsupplyauthority.
SafetyInstructions
1. Usefullfaceshieldprotectionwhenloadingandunloadingthechamber.
2. Fillonlywithhoseorbucket–donotconnecttowatermains.
3. Loadorunloadwithgloveswhichcoverforearmsorwearlongsleeveshirt.
4. If contents are harmful wear an appropriate respirator and filter to prevent inhaling hazardous
vapors.
5. Ensure both hands hold lid handles when opening autoclave and ensure no body parts extend over
theautoclave.
6. Ensurethebasketsarenottooheavywhenloadingandunloadingandusecorrectliftingmethods.
7. Keephandsandbodyawayfromautoclavewheninusetopreventburningand/orscalding.
Operation
1. Connect the chamber to an alternating current supply of voltage specified on the rating plate mounted on
thesideorrearofthechamber.
DANGER:THISAUTOCLAVEMUSTNOTBECONNECTEDTODIRECTCURRENTSUPPLY
1. Ensure the drain 1 is closed, and then manually fill the boiler with clean water to a level just below the top
ofbasketstand.Note:YOUMAYUSEDISTILLEDWATER.
2. Loadtheautoclave.
3. Closethelidfollowingthe“closingthelid”instructions.
4. Switchthemainspoweron.Turnthetemperaturecontrollertotherequiredtemperature.
5. Set the timer to the required sterilisation period. The timer has multiple time autoclave selections which
are displayed in the window i.e. 0.1 seconds, 0.1 minutes, 0.1 hours, hours, 10’s of hours. The time may
beselectedinanymodefrom001to300hrs.
6. Pressthe‘start’button.
7. Theautoclavewillswitchonandwillheatupuntilthepresettemperatureisreached.
8. When the preset temperature is reached the timing cycle will start. The red alarm light will illuminate on
the temperature controller. Constant pressure/temperature will be maintained in the autoclave until the
preset time has elapsed, at which time the heaters will switch off and the dump valve 5 will automatically
releasethesteamifitisopen.
9. Openthelidfollowingthe“openingthelid”instructions.BESUREALLSTEAMISRELEASEDFIRST.
Indicators
1.Anindicatinglampadjacenttothecontrollerwillilluminatewhentheheatersareoperating.
2.Anamberpilotlightwillilluminateifthewaterlevelistoolow.
3. An automatic digital reset timer with multiple time modes will be activated once temperature reaches the
set temperature, the timer will then flash “off” until the time has elapsed, then the autoclave will cool until
the‘start’buttonispressedagain.
4. If you wish to change the temperature you are using the autoclave at and therefore the temperature at
whichthetimerwillbeginitstimingcyclepleasecallthemanufacturerstechnicaldepartment.
5.Itisadvisabletocheckthewaterlevelaftereachcycle.
ChartRecorder(Optionalnotonallmodels)
Pleasereadthechartrecordermanualbeforeuseoftheautoclave.
The recorder is programmed to record the autoclave temperature on Channel 1. If the recorder is remote it
should be mounted/placed adjacent to the autoclave and then using the leads from the recorder, connected
to the matching sockets on the rear of the autoclave body. When positioned and connected to the autoclave,
open the door of the recorder, place the chart paper in position and remove the white caps from the
pens(the internal door can be opened to access the pens), press the button to switch the recorder ON.  
ConnecttherecordertotheMAINSsupplyandturnON.
3LaboratoryEquipmentPtyLtdInstructionManualAutoclaveAA13,16,18,20Ver2.512013