Manual

Operation manual
SEA Series Steam Sterilizers
19
9.5Cleaning Steam generator
9.5.1 Aim: Clean the steam generator and steam pipe, to avoid the
furring block the pipe.
9.5.2 Description: 20 min cleaning + 5 min drying
9.5.3 Purpose: to clean the steam generator furring, take it out of
the machine through pre vacuum.
9.5.4 Operation: when the machine is inactivity, press SELECT for
5 seconds, the light and light will go on at mean time,
it means you chose the cleaning cycle. Close the door and press
START, it will start to clean.
9.5.5 Cleaning remind: after cleaning, every 100 cycles, light
and light will go on at the mean time and glitter. Remind
you to clean the steam generator.
Friendly remind: Cleaning function will not affect the autoclave
cycle, only remind message.
134 121
134
121
Fig 9-10