Manual
Operation manual
SEA Series Steam Sterilizers
15
8.Function setup
8.1 Enter function setup mode:Press STEP key for 20s when on non-work state
time dispaly Ld pressure dispaly 1 enter function setup mode.
8.2 Function mode setup With 7options, press SELECT key, you can select 1-7-1cycle
dispaly. The definition and display of 7 options as following
Option 1 Year dispaly yer press STEP or key increase press DRY
or Key decrease setup numerical value
Option 2 Month dispaly nth press STEP or key increase press DRY
or Key decrease setup numerical value
Option 3 Day display day press STEP or key increase press DRY
or Key decrease setup numerical value
option 4 Hour display hor press STEP or key increase press DRY
or Key decrease setup numerical value
Option 5 Minute display nin press STEP or key increase press DRY
or Key decrease setup numerical value
Option 6 Printing,it displays Pvt ,press STEP key then you can choose the cycle
between on and off . on stand for print , off stand for no printing .
Option 9: Temperature data adjust automatically option, display D-P ,When you setup this
option, you can adjust the temperature more or less than 2C-3C in order to meet your request.
But we suggest user should setup under guidance of technician or not setup.
Caution: If some function setup was changed by pressing the key without intention,please
switch off the power,then restart it.It will exit from the wrong setup.In this case,the function
won't be affected by this.
Option 7 Exit,,when Eit displays,please press START/STOP to save and exit setup
mode.
Option 8 Sterilization time adjust, display S-T when you setup this option, the sterilization
time(SP)win increase at former value. The adjustable range are following 3 kinds. One is
increased by 1 minute, Second is increased by 2 minutes. Third is increased by 3 minutes.
Press SELECT key after setup, you will come back to option 1.At this time, you must press
STEP key 3 seconds in option7.Then you can start the setup of option8 and option. Otherwise
it only can start cycle between option1and option7.