Manual
5.2.4 After the instruments are loaded, you may close
and lock the chamber door by turning the door knob
clockwise to its locking position. The READY
indication light goes on. (Fig 5-4)
Caution: The door knob must be turned to its lock position, otherwise, there might be steam
leak during a cycle and the alarm appears and displays error code E6 . If the chamber is
in keep warm mode, sometimes it is difficult to close the chamber door because there is still
steam inside. You can keep the door open for a while until steam and moisture is vaporized,
and the door is easy to close. You may also push the door inwards with some force while
turning the door knob to its locking position. If there is still steam leak after you did the above,
you may try to readjust the door mechanism by a spanner as per item 6. Door adjustments.
5.2.5 Press START/STOP key, the ready light goes off,
RUNNING light turns on, and display shows HE the
machine starts a sterilization cycle. It takes about 20-45 minutes
to complete an automatic cycle that includes heating up, sterilizing
and drying. The cycle time depends on the volume of the instruments
loaded, the chamber temperature at the beginning of the cycle and
the cycle program you selected.
Fig 5-5
Fig 5-4
Operation manual
SEA Series Steam Sterilizers
9