IN STRUC TION MANUA L Laboratory Equipment Pty Ltd email: sales@labec.com.au Ph: 02 9560 2811 • Fax: 02 9560 6131 www.labec.com.
SEA Series Steam Sterilizers User Registration On the serial number plate at back of the machine, you may find the unit's model number, serial and reference number. Please double check the numbers and fill in the form with and quote the reference number, when you are in contact with your distributors: Model Reference number Serial number * Before operating the sterilizer, please read carefully all safety cautions and instructions for operation.
SEA Series Steam Sterilizers Important information on Safety Stability Smother!!! Heaviness!!! 1.Please check the voltage according to the nameplate before plug the power cord, Don't use electricaloutlets other than the correct voltage. Otherwise it might cause fire or electric shock. 2.Don't wet your hand before plugging or unplugging the power cord. 3.Don't place your sterilizer on an unstable table top, moving desk,tilted and shaking surface 4.
SEA Series Steam Sterilizers Contents 1.Description & Usage .........................................................................1 2.Specifications....................................................................................2 3.Installation.........................................................................................3 4.Control panel............ .........................................................................5 5.Operation......................................................
SEA Series Steam Sterilizers 1.Description & Usage Runyes design and make pressurized steam autoclaves especially to meet the needs for sterilization of instruments and devices in clinics,departments in a hospital and laboratories. Operators include doctors or his/her assistance under doctors supervisions.The autoclave is computerized, user friendly and easy to operate. Its real time display shows all data during the sterilization cycle.
SEA Series Steam Sterilizers 2.Specifications Chamber inner dimensions (Diameter x Depth)..............................................................22L:247mm x 450mm 17L:247mm x 355mm Power supply .........................................................................................................according to the nameplate Sterilizing temperature /pressure..........................121 /(0.9 1.3 Bar(Under standard atmospheric pressure .................................................................
SEA Series Steam Sterilizers 3.Installation The Autoclave should be installed in a drafty place allowing minimum of 10 cm space at all sides and 20 cm at the top. The ventilation openings at the side of the machine should not be blocked. The Autoclave should be placed on a leveled counter top.
SEA Series Steam Sterilizers Preparation before using We installed the draining filter before the device leaving factory. (fig 3-2). Clean the draining filter once per month. You can see the instructions of how to clean it according to page 12. Fig3-2 Before using the device, you should make electrical connections by using the power cord that comes with the unit. Insert the hollowed end of the power cord in to the inlet of the machine and plug the other end on a power outlet.
SEA Series Steam Sterilizers 4.Control panel 1.
SEA Series Steam Sterilizers 5.Operation Please connect the power cord before you start the operation. Press the main power switch at right lower corner of the front panel. The power switch light turns on indicating that electric power is connected. The screen no display, automatically examine. 1 s later, enter to ready to work state after automatically examine. LCD display as fig5-1 shows. RUNYES -Created by love Pls Select a cycle Fig5-1 5.
SEA Series Steam Sterilizers 5.3 START cycle Close the door after loaded the instruments, press START key. 0.5S later, electrical lock will work, with “KA” sounds. If the electrical lock did not achieve expect position will show door fault. It will start sterilization cycle after locked the door successfully. Look at fig 5-4. During sterilization cycle, press MENU/C key for 3S, it will exit cycle.If you want to stop the cycle before drying cycle, it shall need another 3 min to do drying cycle.
SEA Series Steam Sterilizers 5.
SEA Series Steam Sterilizers 6.Printer installation and USB Due to the requirement of customer, Runyes sterilizer can adopt prescriptive printer. Connect the printe to the sterilizer with cables according to fig 6-1. Switch on the power, two lights on the printer panel will goes on, means printer has been installed correctly. According to system setup you can setup whether you need print the process or not. Sterilization Fig6-1 Print information explaining Date:Date Cycle:Cycle Times C.N Cycle name C.S.
SEA Series Steam Sterilizers 7.System setup When the device on the ready to work state, press MENU/C key for 3 S will enter to system setup, see fig 7 -1. It has 4 options for select. Press STEP key and QUICK key to select,Press OK key to enter sub-menu to setup system. DATE TIME PRINTER LANGUAGE Fig7-1 Option1 Date setup With 3 sub-menus (fig 7 - 2). Press STEP and QUICK key to select. Press OK key enter this sub-menu. Press STEP and QUICK key to setup the valve.
SEA Series Steam Sterilizers Option 4: Language setup LANGUAGE SETUP With 2 sub-menus(fig 7-5). Press STEP and QUICK key to select language. Press OK key to save the setup and back to upper level of menu. Press MENU/C key back to upper level of menu directly. CHINESE(S) ENGLISH Fig7-5 Option 5: time setup It is designed to setup or extend the sterilization time, ranging from 0~3 minutes, can meet different needs.
SEA Series Steam Sterilizers 8.Cleaning the draining filter After a long time use, maybe some small impurity will deposit on the draining filter. It will affect vacuum and drying. All the impurity come from the oil or dirt or dust of instrument which need sterilization, also maybe comes from the water. In order to prolong the life of the draining filter, please note the following items. Caution Use eligible distilled water, the instruments should be cleaned before sterilization.
SEA Series Steam Sterilizers 9.Service and Maintainance Clean the water tank every week by using medical disinfectant. First drain the distilled water out, using cotton fabric with medical disinfectant to scour the tank wall, then clean it use distilled water. Clean the sterilizer chamber every month by using alcohol. 9.1 how to clean the water tank. 9.1.1 Please clean the sterilizer water tank and waste water tank regularly. 1 Empty the tanks water, please refer to item 10.2 draining the water.
SEA Series Steam Sterilizers 9.3 Clean the sealing ring regularly After using the sterilizer for a period, the scale and impurity will leave at the surface of sealing cover of the sterilizer, it will affect the performance of sealing. The sealing cover surfaces need to be cleaned. Clean the sealing ring and sealing cover by using cotton fabric with distiller water, see fig 9-5 ,9-6. If the sterilizer leak air after been cleaned, You shall take the sealing ring off, clean it.
SEA Series Steam Sterilizers 10.Transport & Storage 10.1 Preparation for transportation and storage. Power the machine off, unplug the power cord and let the chamber cool down 10.2 Draining water Draining of water from water tank and used water tank. Connect the water tube to the outlet.And un screw the outlet-screw anti-clockwise the outlet-screw, according to (Fit10-1) Used water tank will drain water after turn the draining knob open accordin g to .(Fig 10-2) Fig10-1 Fig10-2 10.
SEA Series Steam Sterilizers 11.Warranty The manufacturer provides a warranty for repair at not change to the customer for two years after the invoice date with presentation of the invoice. Following are the disclaimers: 1 Damage due to improper moving, installing and operating the machine; 2 Damage due to dismantling the machine by unauthorized personal; 3 Unable to present the purchasing invoice or the invoice is altered;.
SEA Series Steam Sterilizers 13.Appendix. Form 1 Work mode chart Unwrappd Work mode chart cycle S cycle 18min S cycle Sterilizing temperature 134 134 134 121 134 Sterilizing pressure 2.10 2.10 2.10 1.10 2.10 S S S S S 0-7min 0-7min 0-7min 0-7min 0-7min Vaeuuning time 4min 4min 4min 4min 4min Admission time 5min 5min 5min 5min 5min Sterilizing time 4min 18min 4min 16min 3.
SEA Series Steam Sterilizers 134 sterilization cycle working chart ( bar) 2.0 Pressure 1.0 0.5 0 Time -1.0 1-2 Pre-heating 5-6 Drying 2-3 Pre-vacuum 6-7 Balance period 3-4 Sterilization 1-7 sterilization 4-5 Exhausting 121 seerilization cycle working chart 2.0 ( Bar) Pressure 1.0 0 Time -1.
SEA Series Steam Sterilizers Pressure sensor Inner temperature sensor Outer temperature sensor Steam generator temperature sensor (RY-B38)Tranformer10v5A (input) (RY-B38)Tranformaer21v (input) (RY-B38)Transformer 10V1.
SEA Series Steam Sterilizers Dressare seasor Exhaust solenoiel valve nornrally open Safety valve Drainging water soleroid valve nornolly open Enters the steam solenoid valve Moisture separator Steam solenoid valve Air filter Chamber Enters the steam pump Vacuum pump Two way three position valve Used water tank Water tank Water pump Draing water connector Fill water connector Yun piping Operation manual 20
SEA Series Steam Sterilizers 14.Trouble shooting 1.During venting or drying, the thump sounds. It means that the Condensed water collector is full. You must drain the water from the collector. 2.During a cycle, a banging sounds and large steam leaks from the chamber door and displays E6 code. Please verify whether the door knob is properly turned and locked. If the steam leaks at the beginning of a cycle, you need to change the seal according to instructions at page 14. 3.
SEA Series Steam Sterilizers 15.Important information 1.The autoclave must be installed on a leveled counter top. 2.When operated at a higher elevation of more than 500 meters,the machine must be reprogrammed. Please get in touch with your distributor for instructions. 3.Distilled water must be used for prolonging the life of the machine. 4.The radiation openings in the machine must not be blocked or covered. 5.