User Manual
SEA Series Steam Sterilizers
Operation manual
9
Fig 5-4
5.2.5 After the instruments are loaded, you may close andlock
the chamber door by turning the door knob clockwiseto its
locking position. The ready indication light goeson. (Fig 5-
4)
Caution: The door knob must be turned to its lock position,
otherwise, there might be steam leak during a cycle and the
alarm goes on and displays error code E6 . If the chamber
is in keep warm mode, it is sometimes difficult to close the
chamber door because there is still steam inside. You can keep
the door open for a while until steam and moisture is vaporized,
and the door is easy to close. You may also push the door inwards
with some force while turning the door knob to its locking position. If there is still steam leak
after you did the above, you may try to read just the door mechanism by a spanner as per item
6. Door adjustments.
5.2.6 Press START/STOP key, the ready light goesoff
and display HE the machine starts a sterilization cycle.It takes
about 30-60 minutes to complete an automatic cyclethat includes
heating up, sterilizing and drying. The cycle timedepends on the
volume of the instruments loaded, the chamber
temperature at the beginning of the cycle and the cycle program
you selected.
5.2.7 When a cycle completes, the display Ed the indication light
RUNNING Turns off and the indication light READY turnson
with sounds. At this time you may then open the chamber door andunload
the instruments which are sterilized. (Fig 5-6) At mean time,temperature
indicate prt , it means that it is performing print,whether you have install
printer, the process can not be omitted. But youneedn't care about it. It will
not effect the process of opening the doorand unloading the instruments.
The indication will disappear whenprinting is over.
Fig 5-6
Fig 5-5