Manual

Maintenance
53
How to Sanitize the Tank only
A tank sanitization is performed to eliminate bacteria growth in the tank only. It is recommended to sanitize the
tank at least once a year.
Things to Know BEFORE you sanitize the Tank
HAZARD
All safety precautions must be followed when handling Hydrogen Peroxide.
Rubber gloves, safety goggles and a lab coat must be worn to avoid any skin
and body contact.
A
If you have performed a system sanitization, then you do not need
to perform a tank sanitization.
You will need a minimum Tank Level display of at least 50% (A).
B
Locate the clear elbow fitting, the clear tubing and the syringe in
the Sanitization Kit (B).
You will need 200 ml of 30% Hydrogen Peroxide solution and
200 ml of purified water. Hydrogen Peroxide is readily available
through most Scientific Supply Companies.
The Vent Filter will need to be replaced after the sanitization is
completed.
The system will not be able to deliver Product Water while the
system is being sanitized.
ATTENTION The total time needed to sanitize the tank is at least 4.5 hours.