User guide
Table Of Contents

Troubleshooting
37
How to Replace the Filter Pack Cartridge
ATTENTION
Open the Tank Outlet Valve before installing a new Filter Pack cartridge. This
keeps the tank from filling until the Filter Pack cartridge is rinsed out (A).
HAZARD
Do not touch the UV Lamp when replacing the Filter Pack cartridge.
Removing the Filter Pack cartridge
A
Open the front cover (this may require removal of the final filter) to go
into STANDBY mode. Wait for the system to depressurize.
The Display
will show STANDBY blinking for 10
seconds.
•
Remove the Final Filter.
B
•
Press your thumbs on the system and Swing the pack towards you
(B).
C
Remove the pack from the system (C).