User Manual

66
INSP/TESTS No. CONDITIONS AND REQUIREMENTS INST/LCC# RESULTS PASS
partially wetted. Top of heater must be partially unwetted.
Close door and resume DIAGNOSTICS
WASH PUMP Visual
DETERGENT CUP Audible
HEATER 7.8 - 9.6 Amp Ammeter
Return to TAP – Allow water to fill washer until it turns off. Press
and hold RUN – Water must turn off before it overflows.
Visual
DRAIN / COOL – Allow all water to drain Visual
DRY BLOWER Audible
LIQ DETERG PUMP .1 - .2 Amp Ammeter
LIGHT Visual
FAN Audible
RINSE AID PUMP .1 - .2 Amp Ammeter
Run Test
1 Close Detergent Cup Door
2 Scroll through Programs
Confirm Display Information Visual
3 Place known amount of water in test dispense bottles
4 Select Glass Plus
Wash 2 Temp 93°C
Wash 2 Time 10 min
Steam Yes
No. Rinses 3
No. Pure Rinses 1
Final Rinse Temp 60°C
Dry Temp 60°C
Dry Time 20 minutes
Detergent Dispense 20 ml
Rinse Aid Dispense 6 ml
Start Data Acquisition Computer – Identify file with washer serial
number and date
Press RUN
Confirm operation of washer, temperatures and times
Acceptance Criteria:
Visual-
RS232
Wash 2 and Final Rinse Temp +/- 3°C
Dry Temp +/- 8°C once reached