User Manual
Labconco Glassware Washer Service/Technical Manual 
45 
7.  Use diagnostics to check fill times and function of the new valve. 
8.  Re-install the panels. 
DI/Pure Water Pump Replacement 
Use the following procedures to replace/install the DI/Pure Water 
Pump. This pump allows the user to supply the washer with non-
pressurized, gravity-fed purified water. The pump works in 
conjunction with the DI/Pure Water solenoid valve to fill the washer 
when set to use Purified water (during rinses). If the pump fails, 
the washer will experience a Low Water Alarm during DI Rinse cycles. 
**Completely drain and disconnect electrical power to the unit before 
proceeding with any of these instructions. 
1.  Remove the Left Panel or Left Service Access Panel. 
2.  Locate the DI/Pure Water Pump. 
3.  Remove the four nuts holding the pump to the Base Weldment. 
4.  Loosen the clamps on the pump inlet and outlet hoses. 
5.  Remove the hoses and place them in their appropriate fittings onto the 
new pump. 
6.  Tighten the clamps on the pump inlet and outlet. 
7.  When the Hoses are secured to the pump, place it over the four mounting 
screws. 
8.  Tighten down the four lock-nuts. 
9.  Test the pump in diagnostics. 
Water Level Pressure Switch/Hose Replacement 
Use the following procedures to replace/install the DI/Pure Water 
Pump. This pressure switch has replaced the domed float switch in 
older units. As the water fills the sump/tank the pressure in the 
tube rises actuating the first contact (initial fill). If the second 
Figure 18 
Left: DI/Pure Water Pump. The 
Pure water pump is located in the 
Front Left of the machines base. 
The Yellow Circles indicate the four 
nuts that mount the DI Pump to the 
base of the unit. 
The Blue and the Green Circles 
highlight the DI Pump’s inlet and 
outlet hose fittings and clamps, 
respectively. The pump’s inlet hose 
leaves the inlet solenoid and loops 
over the pump. The pump’s outlet 
hose leaves the pump and goes to the 
tank’s Water Inlet Port. The location 
of this port depends on the age of the 
unit; it is either on the back or the 
right side of the unit. 
The Magenta Circle points out the 
electrical connections where the DI 
Water Pump’s wiring hooks into the 
units Base Wiring Harness. 