Appendix C: RapidVap Specifications
Product Service Domestic 1-800-522-7658, International 816-333-8811
Evaporation Rates
RapidVap Vacuum Models
The following tables show the approximate time to
evaporate various solvents in 75mm and 16mm sample
tubes when the sample block is fully loaded. Other sample
blocks are available for different size tubes. If they are
used or if a different number of samples is run at one time,
the evaporation times can be estimated from these charts or
determined experimentally.
Evaporation Time – 75 mm dia. Sample Tube
(Time in Minutes)
Solvent:
Methylene
Chloride
Toluene
Acetonitrile
Water
Block Preheat &
Block Set Point
Temperature:
50°C
80°C
75°C
80°C
Vacuum:
250 mbar 133 mbar 167 mbar 133 mbar
Starting Volume
300ml 62 90 102 402
250 56 72 84 345
200 49 66 71 289
150 42 57 60 239
100 36 33 49 189
50 22 40 37 95
Number of samples: 8
Vortex speed: 28%
Vacuum pump: Labconco model – 79249-00
Rated free air flow – 88 L/min
Max. attainable vacuum – 31 mbar
V