Refer to the following if your RapidVap fails to operate
properly. If the suggested corrective actions do not
solve your problem, contact Labconco for additional
assistance. The following failure codes may appear on
the display when problems are sensed by the internal
self check routine.
DISPLAY
ERROR CODE
CAUSE CORRECTIVE ACTION
Top Temp Sensor
Top sensor failure Replace top sensor assembly.
Connection failure Repair connection.
Bottom Temp
Sensor
Bottom sensor
failure
Replace bottom sensor assembly.
Connection failure Repair connection.
Vacuum Sensor
Sensor failure Replace sensor assembly.
Connection failure Repair connection.
Lid Opened
Lid open Close lid.
Switch failure Replace switch.
Connection failure Repair connection.