Chapter 6: Maintaining Your Clean Bench
Weekly
Wipe down the interior surfaces of the clean
bench with a mild household detergent or
cleaner and allow to dry.
Using a damp cloth, clean the exterior surfaces
of the clean bench, particularly the front and top
of the clean bench to remove any accumulated
dust.
Monthly (or more often as required)
Check all service valves, if so equipped, for
proper operation.
Check the prefilters and replace if necessary.
The prefilters should be replaced at least
quarterly.
All weekly activities.
Quarterly
Replace the prefilter. See page 57 for ordering
information.
All monthly activities.
Annually
Have the clean bench recertified by a qualified
certification technician.
All quarterly activities.
Biannually
Replace the fluorescent lamp if required.
Replace the UV lamp (if included).
All annual activities.
Service Operations
Resetting the Circuit Breaker:
1. The circuit breaker is located next to the power cord
on the upper rear panel of the bench as shown in
Figure 6-1. If the circuit breaker trips, it can be
reset by pressing the white button in.