Chapter 5: Using Your Clean Bench
Wipe down the interior surfaces of the clean
bench with a mild household detergent. DO
NOT use abrasive cleaners, bleach or solvents,
as they may damage the work surface of the
clean bench.
Allow the clean bench to operate undisturbed
for 5 to 15 minutes before loading materials.
Wear long sleeved lab coat with knit cuffs and
over-the-cuff rubber gloves. Use protective
eyewear.
Loading Materials and Equipment
Only load the materials required for the
procedure. Do not overload the clean bench.
Do not obstruct the air diffuser.
Large objects should not be placed close
together.
After loading the clean bench, wait 2 to 3
minutes to purge airborne contaminants from
the work area.
Work Techniques
Keep all materials at least 4 inches inside of the
clean bench and perform all contaminated
operations as far to the front of the work area as
possible.
Segregate all clean and contaminated materials
in the work area.
Arrange materials to minimize the movement of
contaminated materials into clean areas.
Keep all discarded contaminated material to the
front of the clean bench.
Avoid moving materials or the operator’s hands
and arms in and out of the work area during use.
Avoid the use of an open flame.
Use proper aseptic technique.
Avoid using techniques or procedures that
disrupt the airflow patterns of the clean bench.
Final Purging
Upon completion of work, the clean bench
should be allowed to operate for 2 to 3 minutes
undisturbed, to purge airborne contaminants
from the work area.