Chapter 5: Using Your PuriCare
Working In Your PuriCare
Planning
Thoroughly understand procedures and
equipment required before beginning work.
Arrange for minimal disruptions, such as room
traffic or entry into the room while the station is
in use.
Start-up
Turn on fluorescent light and station blower.
Check the return air grilles for obstructions, and
note the pressure gauge reading.
Allow the station to operate unobstructed for 5
minutes.
Wash hands and arms thoroughly with
germicidal soap.
Wear a long sleeved lab coat with knit cuffs and
over-the- cuff rubber gloves. Use protective
eyewear. Wear a protective mask if appropriate.
Wipe-Down
Pivot the sash to its full open position.
Wipe down the interior surfaces of the station
with 70% ethanol, or a suitable disinfectant, and
allow to dry.
Slowly close the sash.
Figure 5-2